Medical-Elektro.pl

Medical ventilator Evita

Producer: DRAGER
Model: EVITA
Currently available:NO
Technical condition: Very good
Price 3.900,- PLN 893 € gross

The prices in PLN and EUR is not a binding offer in accordance with the regulations of the Polish Civil Code and are given for informational purposes only. To obtain a quote, please contact our sales department (more informations).

The prices in EUR depend on the exchange rate and are updated daily from Monday to Friday according to the average exchange rate of the Polish National Bank (Source: http://www.nbp.pl/home.aspx?f=/kursy/kursya.html). The binding price of our products in Euro is the price according to the exchange rate from day of order.

Equipment

Przewody podłączeniowe

Aparat na statywie jezdnym

Description

Respiorator Draeger EVITA.

Dane techniczne (z braku danych technicznych w języku polskim, prezentujemy Państwu dane techniczne w j. angielskim)

Ventilation Settings:

Ventilation mode

  - IPPV, IPPVAssist (CMV, CMVAssist)

  - SIMV, SIMVASB (SIMV, SIMV/Psupp)

  - MMV, MMVASB (MMV, MMV/Psupp)

  - BIPAP1), BIPAP1)ASB, BIPAP1)Assist
    (PCV+, PCV+/Psupp, PCV + Assist)

  - CPAP, CPAPASB (CPAP, CPAP/Psupp)

  - APRV (optional)

  - ILV (optional)

Enhancements

  - AutoFlowTM – Automatic adaptation of
    inspiratory flow in volume controlled modes

  - ATCTM – Automatic Tube Compensation (optional)

  - NIV - Mask Ventilation (optional)

Ventilation frequency (f): 0 to 100 /min, 0 to150 /min (Neonatal)

Inspiration time (Tinsp): 0.1 to 10 s

Tidal volume (VT) (BTPS):

  - 0.1 to 2.0 L (Adult)

  - 0.02 to 0.3 L (Pediatric)

  - 0.003 to 0.1 L (Neonatal)

Inspiratory flow:

  - 6 to 120 L/min (Adult)
  - 6 to 30 L/min (Pediatric and Neonatal)

Inspiratory pressure: 0 to 80 mbar (cmH2O)

PEEP / intermittent PEEP: 0 to 35 mbar (cmH2O)

Pressureassist (PASB) (Psupp): 0 to 80 mbar (cmH2O)

Rise time for inspiratory pressure: 0 to 2 s

O2 concentration: 21 to 100 Vol.%

Multi-sense Trigger Criteria: Internal automatic pressure trigger, Flow, Volume (Flow adjustable 0.3 to 15 L/min)

Measured values displayed:

Airway pressure: Peak pressure, plateau pressure, mean pressure, PEEP, min. pressure (0 to 99 mbar/cmH2O)

Minute volume (MV), (BTPS): MV, MVspont, MVleak (0 to 99 L/min)

Tidal volume (VT), (BTPS): Inspired VT, expired VT, VTASB (VTPS) (0 to 3999 mL)

Breathing frequency (f): ftotal, fspon, fmand. (0 to 300/bpm)

O2 concentration: (FiO2) Inspired O2 concentration (15 to 100 Vol.%)

Lung mechanics:

  - Resistance ((0.0 to 600 mbar/L/s)(cmH2O/L/s)

  - Compliance ((0.0 to 300 mL/mbar)(mL/cmH2O)

Breathing gas temperature: 18°to 51°C

Waveforms: Airway pressure-time, flow-time, volume-time, ...

Trends (optional): FiO2, MV, VT, f, PEEPi, R, C, etCO2, ...

Loops (optional): Paw-V, V-Flow, Flow-Paw, ...

Capnography (etCO2) (optional): 0 to 100 mmHg

CO2 production (VCO2): 0 to 999 mL/min, STPD

Serial dead space Vds: 0 to 999 mL, BTPS

Dead space ventilation (Vds/VT): 0 to 99%

Alarms / Monitoring:

Airway pressure: High/Low

Expired minute volume: High/Low

Tidal volume: High

Apnea alarm time: 5 to 60 s

Spontaneous breath frequency: High

Inspired O2 concentration: High/Low

Breathing gas temperature: High

SpO2 pulse (optional): High/Low

etCO2 (optional): High/Low

Performance data:

Max. flow for pressure support and spontaneous breathing: 180 L/min (adult), 60 L/min (pediatric)

Valve response time: ≤ 5 ms

Control principle: Time cycled, volume constant, pressure-controlled

Safety relief valve: 100 mbar (cmH2O)

Leakage compensation

Hose system compensation

Outlet for pneumatic nebulizer

Operating data:

Mains power connection: 100 to 240 V, 50/60 Hz, 10 to 30 V DC (optional)

Power consumption: Approx. 125 W

Gas supply operating pressure: O2, air: 2.7 to 6 bar/39 to 87 PSI

Physical Specifications:

Dimensions ventilator (WxHxD): 530x290x450 cm (20.9x11.4x17.7 inches) (without trolley)

Weight basic unit: Approx. 27 kg (60 lbs.)

Diagonal screen size: 6.5" TFT LCD color display

Machine outputs:

Digital output: Output and reception via an RS 232 C interface

Digital output (optional): For output and reception via two RS 232 C interfaces

Analog output (optional): For analog output of two measured values

Gallery

To see the pictures at higher resolution, click the image below.